LDP5 Meter Setup and Operation Instructions
312668W 25
Error Codes
Error codes are listed below. Even in an error condition,
the unit keeps track of the amount dispensed. With any
error code displayed, as shown at right, you can:
Press the Manual/Reset
button. Error code is
cleared, unit switches into
Manual mode, and dis-
pensed amount is dis-
played.
Press the Auto/Reset but-
ton. Error code is cleared,
unit switches into Auto
mode, and the preset
amount is displayed.
FIG. 33
GAL
QTSPTS
LITERS
R-TOTAL
AUTO
ti12061a
Error Code Cause Solution
Err 1 Flow rate is higher than 5 gpm. Adjust the flow rate so it is not higher
than 5 gpm.
Air was pumped through the line. Purge air from the line.
Err 4 Flow has continued after it should
have shut off.
Check if unit is resting on the trigger if
an obstruction is pressing the trigger.
The unit checks for flow every sec-
ond and repeats the error code until
the trigger is released and the error
code cleared.
Err 5 The unit has dispensed the shut-off
default amount and has stopped fluid
flow.
Press the Manual/Reset button, and
dispense again. To change the
shut-off default amount, see Shut-off
Default Amount, page 21.
Err 6 A preset dispense amount of zero
was entered for the dispense or is
stored as the default, and a Preset
dispense was attempted.
Enter an amount that is not zero. See
Dispensing Fluid in Auto Mode,
page 23.